Narishige Pc10 Manual New Jun 2026

The secondary heating phase triggers automatically, followed immediately by the final drop that breaks the glass into two identical pipettes. Carefully remove both pipettes from the clamps. 5. Parameter Optimization and Troubleshooting

Raise the upper slider until it locks into the top position.

Press Start. If the glass does not melt within 15 seconds, turn off the unit, increase the dial by 5 units, and try again with a fresh capillary.

Controls the electrical current sent to the platinum coil during the first and second pull stages. Higher numbers increase the temperature. narishige pc10 manual new

: A signature acrylic shroud shielded the glass capillaries from external air drafts, preventing even tiny temperature fluctuations from ruining a pull.

(or equivalent 1.0mm O.D.) glass capillary through the heater coil.

This is the essential heating element for the PC-10 (and the PC-100). Its technical specifications are as follows: Controls the electrical current sent to the platinum

Controls the overall length and "shoulder" shape of the pipette.

Slide the glass capillary through the center of the heating coil. Center the capillary relative to the coil. Tighten the upper clamping knob securely.

Rotate the bottom switch 90 degrees clockwise to clear the weight path. Start Switch again. The unit will heat at the NO. 2 HEATER level and pull the glass until it separates. NARISHIGE WEB NEWS Reference Settings Examples Application First Pull Distance Thin/Long Pipette 3 (2L, 1H) Firm/Stubby Tip 4 (2L, 2H) Safety & Maintenance Tips No.051 Setting Values in the PC-10 Puller (Revised) including better illustrations

: Digital displays show heater output (0–100), where 100 represents approximately 2.5V, enabling repeatable mass production.

If you need to source a replacement heater for your PC-10 or its successor, the PC-100, the process is straightforward. You can order the necessary consumables and parts through specialized scientific equipment suppliers.

: New manuals or updates may provide refined operational guidelines, including better illustrations, more detailed step-by-step procedures, and additional tips for optimizing performance.